A Fast Liquid Chromatography Tandem Mass Spectrometric Analysis of PETN (Pentaerythritol Tetranitrate), RDX (3,5-Trinitro-1,3,5-triazacyclohexane) and HMX (Octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine) in Soil, Utilizing a Simple Ultrasonic-Assisted Extraction with Minimum Solvent
Date
2016Author
Aydın, Muhammet
Apak, Resat
Avci, Gulfidan Yenel
Cengiz, Salih
Anilanmert, Beni
Metadata
Show full item recordAbstract
Direct analyses of explosives in soil using liquid chromatography tandem mass spectrometry (LC-MS/MS) methods are very limited in the literature and require complex procedures or relatively high amount of solvent. A simple and rapid method was developed for the determination of pentaerythritol tetranitrate (PETN), 3,5-trinitro-1,3,5-triazacyclohexane (RDX) and octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX), which are among the explosives used in terrorist attacks. A one-step extraction method for 1.00 g soil with 2.00 mL acetonitrile, and a 8-min LC-MS/MS method was developed. The detection limits for PETN, RDX and HMX were 5.2, 8.5 and 3.4 ng/g and quantitation limits were 10.0, 24.5, 6.0 ng/g. The intermediate precisions and Horwitz Ratio's were between 4.10 - 13.26% and 0.24 - 0.98, in order. This method was applied to a model post-blast debris collected from an artificial explosion and real samples collected after a terrorist attack in Istanbul. The method is easy and fast and requires less solvent use than other methods.
Collections
- Makale [92796]