dc.identifier.citationFilik H., Avan A. A. , Aydar S., Cetintas G., "Determination of Acetaminophen in the Presence of Ascorbic Acid Using a Glassy Carbon Electrode Modified with Poly(Caffeic acid)", INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, cilt.9, sa.1, ss.148-160, 2014
dc.identifier.issn1452-3981
dc.identifier.othervv_1032021
dc.identifier.otherav_691fa4f1-82e5-4417-bf3e-a970f76c755a
dc.identifier.urihttp://hdl.handle.net/20.500.12627/72856
dc.description.abstractA stable electroactive thin film of poly(caffeic acid) was deposited on the surface of a glassy carbon electrode by potentiostatic technique in an aqueous solution containing caffeic acid. The electrochemical behaviors of acetaminophen on poly(caffeic acid) modified glassy carbon electrodes (poly(CA)-GCEs) were investigated by cyclic voltammetry and square-wave voltammetry (SWV). The results obtained that the poly(CA)-modified electrode was exhibited excellent electrochemical activity on the oxidation of acetaminophen. The calibration curve for APAP was shown two linear segments: the first linear segment increases from 0.2 to 1.0 and second linear segment increases up to 10 mu M. Using the first range of this calibration plot, a detection limit of 0.026 mu M is obtained for acetaminophen using square wave voltammetry. The poly(CA) electrode is selective not only for the detection of acetaminophen in the presence of ascorbic acid, dopamine, p-aminophenol and uric acid but also selective for the simultaneous determination of these four species present in a mixture. Finally, the proposed method was successfully applied to analysis of acetaminophen in pharmaceutical tablets.
dc.language.isoeng
dc.subjectFizikokimya
dc.subjectElektrokimya
dc.subjectTemel Bilimler
dc.subjectTemel Bilimler (SCI)
dc.subjectKimya
dc.subjectELEKTROKİMYA
dc.titleDetermination of Acetaminophen in the Presence of Ascorbic Acid Using a Glassy Carbon Electrode Modified with Poly(Caffeic acid)
dc.typeMakale
dc.relation.journalINTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE
dc.contributor.departmentİstanbul Üniversitesi , ,
dc.identifier.volume9
dc.identifier.issue1
dc.identifier.startpage148
dc.identifier.endpage160
dc.contributor.firstauthorID15566


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster