dc.contributor.author | Bilge, Hatice Bilge | |
dc.contributor.author | Kesen, Nazmiye Donmez | |
dc.contributor.author | Okutan, Murat | |
dc.contributor.author | Cakir, Aydin | |
dc.date.accessioned | 2021-03-03T19:18:32Z | |
dc.date.available | 2021-03-03T19:18:32Z | |
dc.date.issued | 2015 | |
dc.identifier.citation | Kesen N. D. , Cakir A., Okutan M., Bilge H. B. , "A comparison of TPS and different measurement techniques in small-field electron beams", MEDICAL DOSIMETRY, cilt.40, sa.1, ss.9-15, 2015 | |
dc.identifier.issn | 0958-3947 | |
dc.identifier.other | av_53d62d5f-520c-491a-b0a9-d4739f0f9b6b | |
dc.identifier.other | vv_1032021 | |
dc.identifier.uri | http://hdl.handle.net/20.500.12627/59414 | |
dc.identifier.uri | https://doi.org/10.1016/j.meddos.2014.07.001 | |
dc.description.abstract | In recent years, small-field electron beams have been used for the treatment of superficial lesions, which requires small circular fields. However, when using very small electron fields, some significant dosimetric problems may occur. In this study, dose distributions and outputs of circular fields with dimensions of 5 cm and smaller, for nominal energies of 6, 9, and 15 MeV from the Siemens ONCOR Linac, were measured and compared with data from a treatment planning system using the pencil-beam algorithm in electron beam calculations. All dose distribution measurements were performed using the Gafchromic EBT film; these measurements were compared with data that were obtained from the Computerized Medical Systems (CMS) XiO treatment planning system (TPS), using the gamma-index method in the PTW VeriSoft software program. Output measurements were performed using the Gafchromic EBT film, an Advanced Markus ion chamber, and thermoluminescent dosimetry (TLD). Although the pencil-beam algorithm is used to model electron beams in many clinics, there is no substantial amount of detailed information in the literature about its use. As the field size decreased, the point of maximum dose moved closer to the surface. Output factors were consistent; differences from the values obtained from the TPS were, at maximum, 42% for 6 and 15 MeV and 32% for 9 MeV. When the dose distributions from the TPS were compared with the measurements from the Gafchromic EBT films, it was observed that the results were consistent for 2-cm diameter and larger fields, but the outputs for fields of 1-cm diameter and smaller were not consistent. In CMS XiO TPS, calculated using the pencil-beam algorithm, the dose distributions of electron treatment fields that were created with circular cutout of a 1-cm diameter were not appropriate for patient treatment and the pencil-beam algorithm is not convenient for monitor unit (MU) calculations in electron dosimetry. (C) 2015 American Association of Medical Dosimetrists. | |
dc.language.iso | eng | |
dc.subject | Onkoloji | |
dc.subject | Nükleer Tıp | |
dc.subject | Sağlık Bilimleri | |
dc.subject | İç Hastalıkları | |
dc.subject | Dahili Tıp Bilimleri | |
dc.subject | Tıp | |
dc.subject | RADYOLOJİ, NÜKLEER TIP ve MEDİKAL GÖRÜNTÜLEME | |
dc.subject | Klinik Tıp (MED) | |
dc.subject | Klinik Tıp | |
dc.subject | ONKOLOJİ | |
dc.title | A comparison of TPS and different measurement techniques in small-field electron beams | |
dc.type | Makale | |
dc.relation.journal | MEDICAL DOSIMETRY | |
dc.contributor.department | İstanbul Üniversitesi , , | |
dc.identifier.volume | 40 | |
dc.identifier.issue | 1 | |
dc.identifier.startpage | 9 | |
dc.identifier.endpage | 15 | |
dc.contributor.firstauthorID | 85557 | |