dc.contributor.author | Ahmed, Asm Sabbir | |
dc.contributor.author | DEMİR, Mustafa | |
dc.contributor.author | Gurmen, Tevfi K. | |
dc.contributor.author | DEMİR, Bayram | |
dc.contributor.author | Babalik, Erhan | |
dc.date.accessioned | 2022-02-18T10:59:38Z | |
dc.date.available | 2022-02-18T10:59:38Z | |
dc.date.issued | 2008 | |
dc.identifier.citation | DEMİR B., Ahmed A. S. , Babalik E., DEMİR M., Gurmen T. K. , "Verification and uniformity control of doses for Sr-90/ Y-90 intravascular brachytherapy sources using radiochromic film dosimetry", JOURNAL OF MEDICAL PHYSICS, cilt.33, sa.2, ss.54-59, 2008 | |
dc.identifier.other | av_cd3bbe79-2935-4e81-a315-3f6eac412dbb | |
dc.identifier.other | vv_1032021 | |
dc.identifier.uri | http://hdl.handle.net/20.500.12627/180296 | |
dc.identifier.uri | https://doi.org/10.4103/0971-6203.41193 | |
dc.description.abstract | Intravascular brachytherapy (IVBT) is a useful treatment modality for the recurrence of in-stent restenosis following drug-eluting stents (DES) or IVBT failure. The objective of this study was to measure the dose rate of Sr-90/Y-90 IVBT sources for comparison with that given by the manufacturer and to control the dose uniformities of these sources along the source axis. The dose rates of Sr-90/Y-90 beta sources were measured with a radiochromic film in a custom-made phantom. The films for calibration were irradiated using Co-60 photon beams. The results for the three sources were 4.5%, 2.3%, and 3.5% higher than the corresponding certificate values. Maximum and minimum of the dose rates varied within +/- 10% of those at source center; and maximum dose discrepancy for the first Sr-90/Y-90 source train was 8.2%; for the second source train, 7.1%; and for the third source train, 5.1%. Our study showed that the dose rates given by the manufacturer for the three Sr-90/Y-90 IVBT sources were reliable and dose uniformities were within +/- 10% along two thirds of the treatment length. | |
dc.language.iso | eng | |
dc.subject | Sağlık Bilimleri | |
dc.subject | RADYOLOJİ, NÜKLEER TIP ve MEDİKAL GÖRÜNTÜLEME | |
dc.subject | Klinik Tıp | |
dc.subject | Klinik Tıp (MED) | |
dc.subject | Tıp | |
dc.subject | Dahili Tıp Bilimleri | |
dc.subject | Nükleer Tıp | |
dc.subject | Radiology, Nuclear Medicine and Imaging | |
dc.subject | Radiological and Ultrasound Technology | |
dc.subject | Health Sciences | |
dc.title | Verification and uniformity control of doses for Sr-90/ Y-90 intravascular brachytherapy sources using radiochromic film dosimetry | |
dc.type | Makale | |
dc.relation.journal | JOURNAL OF MEDICAL PHYSICS | |
dc.contributor.department | İstanbul Teknik Üniversitesi , Fen Fakültesi , Fizik Bölümü | |
dc.identifier.volume | 33 | |
dc.identifier.issue | 2 | |
dc.identifier.startpage | 54 | |
dc.identifier.endpage | 59 | |
dc.contributor.firstauthorID | 3375736 | |