dc.identifier.citationGul M., Erkanli K., Erol M. K. , BAKIR İ., "Ventricular Embolization of Edwards SAPIEN Prosthesis Following Transcatheter Aortic Valve Implantation", JOURNAL OF INVASIVE CARDIOLOGY, cilt.24, sa.10, ss.537-538, 2012
dc.identifier.issn1042-3931
dc.identifier.othervv_1032021
dc.identifier.otherav_b775df02-ddcf-4c5d-b0ef-d23eee92b4b6
dc.identifier.urihttp://hdl.handle.net/20.500.12627/179809
dc.description.abstractTranscatheter aortic valve implantation (TAVI) is an alternative therapy in patients with severe aortic stenosis (AS) and high surgical risk. Despite continuous improvements in operators' expertise and device technology, complications associated with TAVI are common. We present a case in which an Edwards SAPIEN prosthetic valve dislocated to the left ventricular outflow tract with hemodynamic collapse 4 hours following implantation and embolized into the left ventricle (LV) during resuscitation.
dc.language.isoeng
dc.subjectCardiology and Cardiovascular Medicine
dc.subjectHealth Sciences
dc.subjectSağlık Bilimleri
dc.subjectKardiyoloji
dc.subjectDahili Tıp Bilimleri
dc.subjectTıp
dc.subjectKlinik Tıp (MED)
dc.subjectKlinik Tıp
dc.subjectCARDIAC ve CARDIOVASCULAR SİSTEMLER
dc.titleVentricular Embolization of Edwards SAPIEN Prosthesis Following Transcatheter Aortic Valve Implantation
dc.typeMakale
dc.relation.journalJOURNAL OF INVASIVE CARDIOLOGY
dc.contributor.departmentIstanbul Mehmet Akif Ersoy Thorac Cardiovasc Surg , ,
dc.identifier.volume24
dc.identifier.issue10
dc.identifier.startpage537
dc.identifier.endpage538
dc.contributor.firstauthorID3380403


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster