Show simple item record

dc.contributor.authorBal, Fatih
dc.contributor.authorOmer, Beyhan
dc.contributor.authorGurdol, Figen
dc.contributor.authorInce, Nurhan
dc.contributor.authorGenc, Sema
dc.contributor.authorAycan-Ustyol, Esra
dc.date.accessioned2021-03-05T20:54:28Z
dc.date.available2021-03-05T20:54:28Z
dc.date.issued2012
dc.identifier.citationGenc S., Omer B., Aycan-Ustyol E., Ince N., Bal F., Gurdol F., "Evaluation of Turbidimetric Inhibition Immunoassay (TINIA) and HPLC Methods for Glycated Haemoglobin Determination", JOURNAL OF CLINICAL LABORATORY ANALYSIS, cilt.26, ss.481-485, 2012
dc.identifier.issn0887-8013
dc.identifier.otherav_d6661b4a-21ba-48cd-8604-ab6be572d6dc
dc.identifier.othervv_1032021
dc.identifier.urihttp://hdl.handle.net/20.500.12627/141486
dc.identifier.urihttps://doi.org/10.1002/jcla.21550
dc.description.abstractBackground Various factors may affect the accuracy of hemoglobin (Hb) A1c measurements that are widely used to monitor glycemic control in diabetic patients. This study was aimed to compare the values of HbA1c obtained by two different methods, Roche Tina-quant second and thirdgeneration HbA1c assays based on the turbidimetric inhibition immunoassay (TINIA), and high-performance liquid chromatography (HPLC) cation-exchange method used by Arkray Adams HA-8160 analyzer. Methods Measurements of HbA1c were carried out in blood samples from 2,917 patients using above-mentioned methods. Linear regression was used for the correlation analysis and linear equations. BlandAltman plots were performed from method comparison data using MedCalc statistical software. Results For the low control, the second generation Tina-quant assay had within-run and between-run CVs 0.8% and 0.9%; for the high control within-run and between-run CVs were 1% and 0.96%, respectively. HPLC method for the low control had within-run CV 1% and between-run CV 1.3%; for the high control within-run CV was 0.6% and between-run CV was 0.9%. Conclusion There was a good concordance between the results of TINIA and HPLC methods (y = 1.091x 0.363; r(2) = 0.96). J. Clin. Lab. Anal. 26:481-485, 2012. (C) 2012 Wiley Periodicals, Inc.
dc.language.isoeng
dc.subjectKlinik Tıp (MED)
dc.subjectSağlık Bilimleri
dc.subjectTIBBİ LABORATUVAR TEKNOLOJİSİ
dc.subjectKlinik Tıp
dc.subjectTıp
dc.titleEvaluation of Turbidimetric Inhibition Immunoassay (TINIA) and HPLC Methods for Glycated Haemoglobin Determination
dc.typeMakale
dc.relation.journalJOURNAL OF CLINICAL LABORATORY ANALYSIS
dc.contributor.departmentİstanbul Üniversitesi , ,
dc.identifier.volume26
dc.identifier.issue6
dc.identifier.startpage481
dc.identifier.endpage485
dc.contributor.firstauthorID17324


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)

Show simple item record